Single/ double/ triple pass option, touch screen, automatic and manual operation, additional raw water tanks, automatic cleaning&disinfection, timed switch on/off, DOW membrane, copper free, night/holiday standby mode.Capacity can be modified based on demand.

Main Features

1.R.O. Water Purification Machine Series

-Meets or exceeds U.S. AAMI dialysis water standard and U.S.ASAIO dialysis water requirement.
-Automatic and manual operation.
-Automatic rinse cycle during standby mode.
-Additional raw water tanks for reliable operation.
-Double pass R.O. ( ultra-pure ) product water to meet HDF water quality requirements.
-Automatic cleaning, disinfection and pure water recycling functions during standby mode to minimize bacteria growth.
-Seamless R.O. casing to minimize the dead space within the reverse osmosis membranes.
-High quality reverse osmosis membrane, high pressure pumps, UV sterilizers, controllers and other assembly parts.
2.System components
-Media filters (with automatic flushing device): remove particulate impurities, manganese ions.
-Activated carbon filter (with automatic flushing device): clear chlorine organics ion.
-Softening filters (with automatic flushing reproducing apparatus): clear calcium and magnesium ion, reducing raw water hardness.
-The reverse osmosis hosts (imported reverse osmosis membrane components): removal ions, bacteria, heat, etc.
-Constant pressure water supply of pure water supply section (full cycle).
-Controller: automatic control system
-Water production (L / H) 25 ℃ | the applicable number of beds.


Specification

Water capacitybased on customer's request
Rated voltageAC 380V/400V/415V/240V, 50/60Hz; 3-Phase 4-wire./(depends on clients' detail situation)
Desalination rate99.8%
Recovery rate65%-85%
Ion removal rate99.5%
Bacteria & endotoxin remove rate 99.8%
Working temperature5-40°C.
Tech adoptedpretreatment + RO system
Pre-treatment Sand filter, Active carbon filter, Water softener
ControlPLC control system adopted