Model: HSP-2008
Product origin: China
Warranty: 1 Year
After-sales Service: Online Technical Support
Model: HSP-2008
Product origin: China
Warranty: 1 Year
After-sales Service: Online Technical Support
The HSP-2008 Infant Incubator is a modern neonatal care device designed to provide a stable, comfortable, and well-controlled environment for newborns, especially premature and low-birth-weight infants. Using microcomputer-controlled servo technology, it precisely regulates air temperature, skin temperature, humidity, and optional oxygen concentration to support critical early-stage neonatal care.
With its quiet operation, excellent visibility, and caregiver-friendly access, the YP-2008 helps medical staff perform daily care and clinical procedures efficiently while minimizing stress and heat loss for the infant. It is well suited for use in NICUs, maternity wards, and delivery rooms.
Main Features
1. Two control modes:Air mode and Baby mode servo controlled by micro-computer
2. The information can be stored,checked,and indicated by the color LCD display screen
3. Humidity and Oxygen concentration display and servo control system
4. “>37°C"temperature set function
5. The air circle with outlet in both front and back of the hood
6. Low noise brushless DC motor
7. Operating doors on three sides
8. X-ray cassette under the bassinet
9. Water reservoir integrated with heater,high temperature disinfection can be applied
10. Inclination of bassinet is adjustable
11. The height of whole unit can be adjusted
12. Multiple failure alarm indications
13. RS-232 connector for data output
Specification
| Power supply | AC220V-230V/50Hz or AC110-120V/50-60Hz, 1300VA |
| Control Mode | Air mode and Baby mode controlled by micro -computer |
| Air temperature control range | 25°C-37°C(override mode 37°C-39°C) |
| Baby temperature control range | 34°C-37°C (override mode 37°C-38°C) |
| Accuracy of skin temperature sensor | ±0.2°C |
| The temperature fluctuation range between incubator temperature and average incubator temperature (under stable temperature condition) | ≤0.5°C |
| Temperature uniformity | ≤0.8°C(mattress is on the horizontal position)≤1.0°C(mattress is on the maximun tilt position) |
| Inclination of the bassinet | ± 12° |
| Noise inside hood | ≤45dB(A) |
| Failure alarm | over-temperature alarm, temperature deviation alarm, sensor alarm , motor fan failed alarm, power failure alarm, water tank position alarm, system alarm and so on. |
| Humidity display range | 0%RH~99%RH |
| Humidity control range | 0%RH~-90%RH |
| Accuracy of humidity display | ±5%RH |
| Accuracy of humidity control | ± 5%RH |
| Indication range of oxygen concentration | 0%-99% |
| Control range of oxygen concentration | 20%~60% |
| Indication accuracy of oxygen concentration | +2% (oxygen concentration setting value is 25% and below) ±3 % (Oxygen concentration setting value is 25% and above) |
| Control accuracy of oxygen concentration | ±4% |
| Weighing Range | 0-8000g (weight system) |
| Resolution for weight indication | 1g (weight system) |
| Accuracy of weigh indication | ± 1% (weight system) |
Need a Turnkey Hospital Solution?