Specification

Physical specifications
Dimensions52*73*123 cm
Weight60 kg
CastersFront wheels with lock
ScreenLCD
Display7 inch
Applications
Patient rangeAdult / Paediatric
LanguageENGLISH
Pipeline supply
Gas configurationO₂, N₂O
Pipeline input range280–600 kPa
Working specifications
PowerAC 100V–240V, 50Hz±2%
Battery Power
Run-timeUp to 2 hrs
Modes of ventilation
VCVVolume Control Ventilation
SIMVSynchronized Intermittent Mandatory Ventilation
MANUALManual
SIPPV (A/C)Synchronized Intermittent Positive Pressure Ventilation
STAND-BYSTAND-BY
Ventilator parameter
Working typeElectronically controlled, pneumatic driven
Tidal Volume range40–1500 ml
Pressure range5–60 cmH₂O
Respiration Rate Range1–100 bpm (1–40 bpm under SIMV)
I:E Range4:1–1:8
Ptrigger Range-20-0 cmH₂O
Minute volume range>18 L/min
SIGH range0.5–100
System Standard
Hypoxic guard systemN₂O cut-off valve, O₂ concentration >25%
Safety valve<12.5 kPa
O₂ flush25–75 L/min
Volume of CO₂ absorber2 L
Flow meter4 tubes, Cascade 0–1.5 L/min O₂; 0–1.5 L/min Air (Optional); 0–1.5 L/min N₂O
Vaporizer1 (with optional for Selectatec), maximum 2 mounts (Isoflurane / Enflurane / Sevoflurane / Halothane)
Monitoring
TypesVT, MV, BPM, Ppeak, Pmean, Inspiratory Platform, FiO₂ (Optional)
WaveformsWaveforms of P-T, F-T
Alarm Limits
Audio/VisualVT too high/too low, MV too high/too low, airway pressure, FiO₂ (Optional), apnea, Oxygen failure, AC power failure, Battery low