Specification
| Physical specifications | |
| Dimensions | 52*73*123 cm |
| Weight | 60 kg |
| Casters | Front wheels with lock |
| Screen | LCD |
| Display | 7 inch |
| Applications | |
| Patient range | Adult / Paediatric |
| Language | ENGLISH |
| Pipeline supply | |
| Gas configuration | O₂, N₂O |
| Pipeline input range | 280–600 kPa |
| Working specifications | |
| Power | AC 100V–240V, 50Hz±2% |
| Battery Power | |
| Run-time | Up to 2 hrs |
| Modes of ventilation | |
| VCV | Volume Control Ventilation |
| SIMV | Synchronized Intermittent Mandatory Ventilation |
| MANUAL | Manual |
| SIPPV (A/C) | Synchronized Intermittent Positive Pressure Ventilation |
| STAND-BY | STAND-BY |
| Ventilator parameter | |
| Working type | Electronically controlled, pneumatic driven |
| Tidal Volume range | 40–1500 ml |
| Pressure range | 5–60 cmH₂O |
| Respiration Rate Range | 1–100 bpm (1–40 bpm under SIMV) |
| I:E Range | 4:1–1:8 |
| Ptrigger Range | -20-0 cmH₂O |
| Minute volume range | >18 L/min |
| SIGH range | 0.5–100 |
| System Standard | |
| Hypoxic guard system | N₂O cut-off valve, O₂ concentration >25% |
| Safety valve | <12.5 kPa |
| O₂ flush | 25–75 L/min |
| Volume of CO₂ absorber | 2 L |
| Flow meter | 4 tubes, Cascade 0–1.5 L/min O₂; 0–1.5 L/min Air (Optional); 0–1.5 L/min N₂O |
| Vaporizer | 1 (with optional for Selectatec), maximum 2 mounts (Isoflurane / Enflurane / Sevoflurane / Halothane) |
| Monitoring | |
| Types | VT, MV, BPM, Ppeak, Pmean, Inspiratory Platform, FiO₂ (Optional) |
| Waveforms | Waveforms of P-T, F-T |
| Alarm Limits | |
| Audio/Visual | VT too high/too low, MV too high/too low, airway pressure, FiO₂ (Optional), apnea, Oxygen failure, AC power failure, Battery low |