Specification
| Technique Specifications | |
| Physical Specifications | |
| Dimensions | 70*85*140cm |
| Weight | 65kg |
| Casters | Front wheels with lock |
| Screen | 12.1 inch LCD touchscreen |
| Working Specifications | |
| Power | AC 220V+/-10%,50Hz+/-2% |
| Pipeline Supply | |
| Gas configuration | O2,N2O,Air |
| Pipeline input range | 280-600Kpa |
| Battery Power | Built-in lithium battery |
| Run-time | ≥180 mins |
| Pre-setting Functions | |
| Language | English, Chinese, etc. |
| Calibration | Automatic |
| Applications | |
| Patient range | Adult/Pediatric |
| Ventilator Specifications | |
| Modes of Ventilation | |
| VCV | Volume Controlled Ventilation |
| SIMV-V | Synchronized Intermittent Mandatory Ventilation-Volume |
| SIMV-P | Synchronized Intermittent Mandatory Ventilation-Pressure |
| MANUAL | Manua |
| A/C | Assist/Control Ventilation |
| PCV | Pressure Control Ventilation |
| PSV | Pressure Control Ventilation |
| STAND-BY | STAND-BY |
| Ventilator Parameter | |
| Working type | Electronic control, pneumatic driven |
| Tidal Volume range | 20-1500ml |
| Pressure range | 5-60cmH2O |
| Rispriation rate range | 1-100bpm(1-40bpm under SIMV) |
| I:E range | 4:1-1:8 |
| Ptrigger range | -20-20cmH2O(based on PEEP) |
| Flow trigger | 0-15L |
| Minute volume range | >18L/Min |
| SIGH range | 0-5/100 |
| System Standard | |
| Hypoxic guard system | N2O cut-off Valve, O2 concentration>25% |
| Safety valve | <12.5Kpa |
| Fresh gas compensation | 25-75L/mi |
| Volume of CO2 absorber | 2L |
| Flow meter | 5tubes,Cascade0.1-15L/minO2;0.1-15L/minNO2,0.1-15L/min air |
| Vaporizer | 2 Selectetac, maximum 2 mounts (Isoflurane/ Enflurane/ Sevoflurane/ Halothane) |
| Monitoring | |
| Display graphics | Waveforms of P-T, F-T,V-T, loops of P-V, V-F, F-P(Zoom in/out, comparison, freeze) |
| Types | VT, MV, BPM, Paw, Lung compliance, Inspiratory Platform, FiO2 |
| Alarm Limits | |
| Audio/Visual | No tidal volume, MV, Paw, FiO2 limitation O2,Air failure, AC power failure, Battery low. |
| PEEP | |
| Type | Integrated, Electronic control |
| Range | OFF, 3-30cmH2O |
| AGSS: Anesthetic Gas Scavenging Systems | |
| Communication Interface: USB, Ethernet, DVl, Interfaces | |